Power Generation Efficiency Characteristics Evaluation System
PEM-2

Inquiries about this product

発電効率特性評価装置 PEM-2イメージ

Inquiries about this product

Property evaluations of thermoelectric generator modules
By applying a maximum 500°C temperature difference across a thermoelectric generator module, this system finds the thermoelectric conversion rate η by finding the power that is generated P when a one-dimensional heat flow Q is passed through the module.

Applications

  • Generation efficiency measurements of thermoelectric generator modules
  • Lifespan testing of modules by thermal cycle testing

Features

  • Uses an Infrared Gold Image Furnace with excellent temperature controllability to allow rapid module performance evaluations and endurance testing.
  • Capable of applying a maximum 500°C temperature difference across the top and bottom surface of the module.
  • Capable of measuring the amount of heat passed through the module.
  • Stably maintains the thermal resistance of the contact surface during heating with the air cylinder mechanism.
  • Measurement behavior can be configured with only software settings by temperature stability judgments, automatic variation of the load on the thermoelectric generator module, and automatic control of the measurement temperature.

Specifications

Measurement value Conversion efficiency, Power generation amount, Penetration heat amount
Measurement method One dimensional heat flow input method
Module size Square 30mm×5 to 30mm t (negotiable)
Heating surface temperature Max. Max. 800 ℃
Sample temperature gradient Approx. 300 ℃ in case of thermal conductivity of Stainless steel of 30 mm t
Pressure on contact surface 2MPa
Atmosphere In inert gas
Required floor space Approx. W3000×D1000 (mm)
Weight Approx. 300 kg
Power source Power : AC 200 V, 1 phase, 4 kVA, One location PC,
Pump : AC 100 V, 1 phase, 1 kVA Outlet two locations
Grounding Resistance 100 Ω max.
one location

Constitution

  • Sample heating assembly(including frame work)  1 set
  • Programmable temperature controller  1 set
  • Measurement circuit assembly  1 set
  • Vacuum evacuation assembly  1 set
  • Data processing unit  1 set
  • Constant temperature water circulator (for cooling lower part heat flow measurement block)  1 set

Utility

Required floor space Approx. W 3000 × D 1000 (mm)
Weight Approx. 300 kg
Power source Power : AC 200 V, 1 phase, 4 kVA, One location
PC, Pump : AC 100 V, 1 phase, 1 kVA
Outlet two locations
Grounding Resistance 100 Ω max.
one location

Typical example of commercially available module

Power evaluation by temperature and load current
Evaluation of temperature and maximum conversion efficiency

Analysis service by using this product

Power Generation Efficiency Characteristics Evaluation System PEM-2 Contact form

When you contact ADVANCE RIKO,

The information that you enter will only be used to provide you with a response. However, when you request that we mail you a catalog or other materials, please understand that the shipper specified by ADVANCE RIKO may use only the information necessary to send those materials, such as your name and address.

Please check our privacy policy for more details.

    RequiredName
    RequiredCompany name
    RequiredPhone number
    RequiredEmail address
    RequiredContents of inquiry