Steady Method Thermal Conductivity Measurement System
GH series

Inquiries about this product

定常法熱伝導率測定装置 GH-1シリーズイメージ

Inquiries about this product

Thermal conductivity evaluations of polymers and glass
A heat-flow method Steady Method Thermal Conductivity Measurement System that complies with the US ASTM E1530 standard.
This system measures comparatively low-thermal conductivity materials over a temperature range of 50 to 280°C.

Applications

  • Thermal conductivity evaluations of semiconductor package materials.
  • Thermal conductivity evaluations of glass substrates.
  • Thermal conductivity evaluations of polymer materials.
  • Thermal conductivity evaluations of ceramic materials.
  • Thermal conductivity evaluations of low-thermal conductivity metals.
  • Thermal conduction measurements of thermoelectric materials.

Features

  • Designed to minimize heat loss in the surface direction with the guard heater.
  • Automatic measurement
    Automatic measurement between 50-300ºC can be possible by only inputting measurement temperature into PC.
  • Accurate measurement
    Calibration data of 304 Stainless Steel, Pyrex, and Vespel have been registered in advance.
    Thermal conductivity of unknown materials can be calculated by the comparison of the calibration data.
  • Sufficient monitor display
    Temperature and thermal conductivity (reference value) of each part of sample holder can be displayed during the measurement.
  • Capable of measuring film (Option).
    Thermal conductivity of relatively low thermal conductivity materials such as thin sheet and film samples can be measured by stacking method.

Specifications

Temperature range 50 ~ 280 °C
Sample size ➀ □ 25 × t 1.5 ~ 8 (mm)
➁ φ 25 × t 1.5 ~ 8 (mm)
➂ □ 50 × t 1.5 ~ 12 (mm)
➃ φ 50 × t 1.5 ~ 12 (mm)
Measurement range Sample size ➀ ➁ 0.1~15Wm-1K-1
Sample siz e➂ ➃ 0.1~20Wm-1K-1
Atmosphere Air

Measurement principle diagram

Calculation formula of thermal resistance

Rs:Thermal resistance of sample
N :Proportional coefficient
TL:Lower side temperature of sample
Tu:Upper side temperature of sample
Q :Heat flux meter output
Ro:Contact thermal resistance of upper and lower interface

By using N calculated from calibration sample, thermal resistance of sample, Rs, is found by the above equation. Thermal conductivity, λ, is given by N substituting Rs into the following equation.

λ:Thermal conductivity of sample
d:Sample thickness

Screen for setting and measurement

*The software is actually Englush version.

Screen for inputting sample conditions.

Screen for measurement.

Screen for displaying date grafh.

A patent and a standard

US standard ASTM E1530

Steady Method Thermal Conductivity Measurement System GH series Contact form

When you contact ADVANCE RIKO,

The information that you enter will only be used to provide you with a response. However, when you request that we mail you a catalog or other materials, please understand that the shipper specified by ADVANCE RIKO may use only the information necessary to send those materials, such as your name and address.

Please check our privacy policy for more details.

    RequiredName
    RequiredCompany name
    RequiredPhone number
    RequiredEmail address
    RequiredContents of inquiry