By applying a maximum 500°C temperature difference across a thermoelectric generator module, this system finds the thermoelectric conversion rate η by finding the power that is generated P when a one-dimensional heat flow Q is passed through the module.
Applications
- Generation efficiency measurements of thermoelectric generator modules
- Lifespan testing of modules by thermal cycle testing
Features
- Uses an Infrared Gold Image Furnace with excellent temperature controllability to allow rapid module performance evaluations and endurance testing.
- Capable of applying a maximum 500°C temperature difference across the top and bottom surface of the module.
- Capable of measuring the amount of heat passed through the module.
- Stably maintains the thermal resistance of the contact surface during heating with the air cylinder mechanism.
- Measurement behavior can be configured with only software settings by temperature stability judgments, automatic variation of the load on the thermoelectric generator module, and automatic control of the measurement temperature.
Specifications
Measurement value | Conversion efficiency, Power generation amount, Penetration heat amount |
---|---|
Measurement method | One dimensional heat flow input method |
Module size | Square 30mm×5 to 30mm t (negotiable) |
Heating surface temperature Max. | Max. 800 ℃ |
Sample temperature gradient | Approx. 300 ℃ in case of thermal conductivity of Stainless steel of 30 mm t |
Pressure on contact surface | 2MPa |
Atmosphere | In inert gas |
Required floor space | Approx. W3000×D1000 (mm) |
Weight | Approx. 300 kg |
Power source | Power : AC 200 V, 1 phase, 4 kVA, One location PC, Pump : AC 100 V, 1 phase, 1 kVA Outlet two locations |
Grounding | Resistance 100 Ω max. one location |
Constitution
- Sample heating assembly(including frame work) 1 set
- Programmable temperature controller 1 set
- Measurement circuit assembly 1 set
- Vacuum evacuation assembly 1 set
- Data processing unit 1 set
- Constant temperature water circulator
(for cooling lower part heat flow measurement block) 1 set
Utility
Required floor space | Approx. W 3000 × D 1000 (mm) |
---|---|
Weight | Approx. 300 kg |
Power source | Power : AC 200 V, 1 phase, 4 kVA, One location PC, Pump : AC 100 V, 1 phase, 1 kVA Outlet two locations |
Grounding | Resistance 100 Ω max. one location |